The Sonel MIC-15K1 insulation resistance meter has been designed to measure insulation resistance of electrical power engineering objects, including single and multi-core cables, transformers, motors and generators, condensers, circuit breakers and other power station equipment. It is especially recommended for measurements in areas with very high electromagnetic disturbances, e.g. at substations with a voltage of 1200 kV AC or 500 kV DC. Due to the measuring voltage of 15 kV (in accordance with the guidelines of ANSI/NETA ATS-2009 TABLE 100.1 standards) the meter can be used for objects with nominal voltage above 34.5 kV.

MIC-15k1 是一款极其现代、符合人体工学且操作直观的设备。它即使在其他仪表将毫无用处的大干扰条件下,也能满足要求最严苛的测量工程师的需求。


功能

  • 测试浪涌保护装置(SPD)。

  • 局部放电指示器PDI.

  • 偏振与去偏振电流(PDC)测量。

  • 基于标准测量值对绝缘系统的诊断 IR、DAR、PI、SV、RT、DD

  • 它允许测试被测对象的电气耐久性 - 击穿电压指示

  • 对电磁干扰具有高抗性,可确保在发电厂以及靠近高压输电线路的 1200 kV 交流 和 800 kV 直流 附近不间断工作。

  • 时间上的平滑电压上升(RT——斜坡测试)能够揭示设备中的局部放电,并对其发生位置进行局部定位。

  • 高效转换器,功率约为150 W,可实现绝缘层烧毁,从而能够通过以下任一方式精确定位电缆和电线的短路位置:

    • 目视检查法(如果整个长度可见电源线),

    • 反射法、地震声波探测器,或使用A型框架来指示断层方向(导体必须埋在地下并与地表直接接触)。

  • 兼容外部软件。

  • 测量结果记忆:990 个单元

  • 单次电池充电的运行时间:约 5 小时

  • 高空作业最高3000米

应用程序

MIC-15k1 兆欧表设计用于测量电力设备的绝缘电阻,即:

  • 单核和多核电缆,

  • 转换器,

  • 电机和发电机,

  • 电站中安装的电容器、开关及其他设备。

特别推荐用于电磁干扰极高的区域的测量,例如带有 1200 kV AC 和 800 kV DC的变电站。得益于15 kV的测量电压(符合ANSI / NETA ATS-2009 TABLES 100.1),该仪表可用于测量额定电压高于34.5 kV的物体。

设备功能

高效高压逆变器,测试电压为15 kV,电流高达10 mA,适用于测量高达40 TΩ的绝缘电阻。取得这样的结果使这些仪表成为无与伦比的设备。采用“GUARD”线进行三线电阻测量,可消除由绝缘污染引起的表面漏电流,从而提高测量结果的可靠性。

该仪表指示介电吸收比 DAR、极化指数 PI以及介电放电 DD的值。

该设备允许用户通过分步(SV - 阶梯电压)或平滑(RampTest - RT)地逐步施加测试电压,来评估绝缘状况。

  • SV方法确保,在良好状态下,介质无论施加何种电压,都能提供相同的结果。

  • RT法可用于确定绝缘材料的特性。该仪器在不使被测物体承受所谓电应力的前提下,平稳地增加测量电压。它记录了绝缘体发生电击穿时的电压值和时间。

内置数字滤波器,平均时间为10、30、60、100、200秒,确保在强电磁干扰区域获得稳定的测量结果

职业倦怠

A very useful solution is the function that allows to Burnout the damaged object. In case of exposed cables, it enables visual identification of the fault location. In the case of shielded cables, the method allows to generate a seismicacoustic wave from the place of damage. In special conditions, an energetic discharge will appear cyclically. By using the geophone it will be possible to precisely pinpoint the place where such a discharge occurs. Burnout feature allows also locating transient faults (appearing, for example, only during rainfall) with the support of reflectometry, and in case of a short circuit (of a screen or return wire) to the ground - applying the method of measuring voltage drop (the A-frame).

Autosaving the measurement results

该设备会自动保存测量结果。自动保存点的数量取决于数据量,这些数据保存在主内存中。

数据分析

Sonel MIC Mobile 移动应用程序允许在测量过程中观察结果。该应用程序可以生成各种配置的实时图表。这使得在测试期间就能评估对象的状况。

测量远程启动和停止功能特别实用。得益于此,测试可以在远程进行,例如在天气条件对用户不利时,可以在另一个房间或车内进行。使用手机GPS,可以精确确定测量位置。

得益于移动应用程序和Sonel Reader软件,用户可以存储先前的测量数据,并将其与从仪表扩展内存中传输的当前结果进行比较。该解决方案可用于准备测量报告、跟踪绝缘劣化进度并规划翻新工程。


绝缘电阻测量

根据 IEC 61557-2 标准的测量范围

UN = 15 000 V: 50 kΩ…40.0 TΩ

直流和增加电压(SV)下的测量ISO 范围分辨率准确性
5 kV0...999 kΩ1 kΩ±(3% m.v. + 10位数字)
5 kV1.00..9.99 MΩ0.01 MΩ±(3% m.v. + 10位数字)
5 kV10.0...99.9 MΩ0.1 MΩ±(3% m.v. + 10位数字)
5 kV100..999 MΩ1 MΩ±(3% m.v. + 10位数字)
5 kV1.00...9.99 GΩ0.01 GΩ±(3% m.v. + 10位数字)
5 kV10.0...99.9 GΩ0.1 GΩ±(3% m.v. + 10位数字)
5 kV100...999 GΩ1 GΩ±(3.5% m.v. + 10位数字)
5 kV1.00...9.99 TΩ0.01 TΩ±(7.5% m.v. + 10位数字)
≥10 kV10.0...20.0 TΩ0.1 TΩ±(12.5% m.v. + 10位数字)
≥15 kV10.0...40.0 TΩ0.1 TΩ±(12.5% m.v. + 10位数字)


“m.v.” - 测量值


取决于测试电压的测量电阻范围
电压 UISO测量范围
50 V200 GΩ
100 伏特400 GΩ
250 V1.00 TΩ
500 V2.00 TΩ
1000 V4.00 TΩ
2500 V10.00 TΩ
5000 V20.0 TΩ
10 000 V40.0 TΩ
15 000 V40.0 TΩ
电容测量
范围分辨率准确性
0...999 nF1 nF±(5% m.v. + 5位数字)
1,00...49,99 μF0,01 μF±(5% m.v. + 5位数字)


  • 显示RISO 测量后的测量容量

  • 对于低于100伏的测量电压,未规定测量误差。